Introduction
7820A GC Safety Manual 11
When handling μECDs:
Never eat, drink, or smoke.
Always wear safety glasses when working with or near open μECDs.
Wear protective clothing such as laboratory jackets, safety glasses, and
gloves, and follow good laboratory practices. Wash hands thoroughly with a
mild nonabrasive cleaner after handling μECDs.
Cap the inlet and outlet fittings when the μECD is not in use.
Connect the μECD exhaust vent to a fume hood or vent it to the outside. See
the latest revision of 10 CFR Part 20 (including Appendix B), or th
e
applicable st
ate regulation. For other countries, consult with th
e
app
ropriate agency for equivalent requirements.
Agilent Technologies recommends a vent line internal diameter of 6 mm
(1/4-inch) or greater. With a line of this diameter, the length is not critical.
WARNING
Do not use solvents to clean the µECD.
WARNING
You may not open the µECD cell unless authorized to do so by your local nuclear
regulatory agency. Do not disturb the four socket-head bolts. These hold the cell
halves together. United States customers removing or disturbing them is a violation
of the terms of the exemption and could create a safety hazard.