EasyManua.ls Logo

Agilent Technologies 7820A

Agilent Technologies 7820A
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 7820A GC Safety Manual
Introduction
Oven thermal leaks
Do not allow wiring or temperature probes to pass through the oven door jam.
Agilent recommends using one of the access holes.
WARNING
The insulation around the inlets, detectors, valve box, and the insulation cups is
made of refractory ceramic fibers. To avoid inhaling fiber particles, we recommend
the following safety procedures: ventilate your work area; wear long sleeves,
gloves, safety glasses, and a disposable dust/mist respirator; dispose of insulation
in a sealed plastic bag; wash your hands with mild soap and cold water after
handling the insulation.
WARNING
Objects passing through the oven door seal can cause thermal leaks which create
hazardous hot spots which cause burns and melt equipment.

Other manuals for Agilent Technologies 7820A

Related product manuals