43
Programming Reference
SCPI Instrument Command List
Command :ARM:SOUR
Long
:ARM[:SEQuence[1]|STARt][:LAYer]:SOURce
Form Set & Query
Parameter
IMMediate|INTernal[1]|INTernal2|EXTernal[1]|MANual
*RST value
IMM
Description Use this command to select the triggering mode of the instrument by
selecting the source of the arming signal:
Use
:ARM:SENSe EDGE | LEVel
to choose between Triggered and
Gated.
Command :DISP
Long
:DISPlay[:WINDow][:STATe]
Form Set & Query
Parameter
ON | OFF | 1 | 0
*RST value
ON
Description Use this command to turn the front panel display on and off. Switching
off the display improves the programming speed of the instrument.
*RST switches the display back on. Use
:SYSTem:PRESet
to perform an
*RST without switching the display back on.
Example To switch off the front panel display:
DISP OFF
Triggering Source :ARM:SOURce Mode
Internal Osc.
PLL
EXT INPUT
MAN key
IMMediate|INTernal[1]
INTernal2
EXTernal1
MANual
Continuous
Triggered | Gated by PLL
Triggered | Gated by EXT IN
Triggered | Gated by MANKey