Programming Reference
SCPI Instrument Command List
Range coupling Amplitude
Description Use this command to program the offset current of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of current, you first have to
execute the
[:SOURce]:HOLD CURRent
command to enable the
:SOURce]:CURRent
subsystem.
The available current range is limited by the combination of:
Specified Voltage limits
Actual OUTPUT Impedance setting
:OUTPut:IMPedance
Actual Expected Load impedance setting
Example To program the offset current of the output signal:
Command :CURR[1]:HIGH
Long
[:SOURce]:CURRent[1][:LEVel][:IMMediate]:HIGH
Form Set & Query
Parameter Numeric
Parameter suffix
A
with engineering prefixes.
Value coupling
Range coupling Low level
*RST value +10 mA (50
into 50
)
Specified Limits 10 V Output (from high Z into short): –396 mA to 400 mA typical
3.8 V (from 50
into short
):
–82 mA to 152 mA typical
Description Use this command to program the high level current of the OUTPUT
signal. Note that to set the OUTPUT levels in terms of current, you first
:HOLD CURR
:CURR:OFF 50MA
Enable CURRENT subsystem
Set OUTPUT offset to 50 mA
Amplitude = High – Low
Offset =
High – Low
2