Programming Reference
SCPI Instrument Command List
Example To program the pulse width by means of the delay parameters:
Command :PULS:TRAN[1]:HOLD
Long
[:SOURce]:PULSe:TRANsition[1]:HOLD
Form Set & Query
Parameter
TIME | WRATio
*RST value
TIME
Description Use this command to set the coupling between transition times and the
pulse width:
Example To set the coupling between transition times and the pulse width:
:PULS:DEL 500NS
:PULS:DEL:HOLD TIME
:PULS:TDEL 750NS
Set OUTPUT delay to 500 ns
Hold OUTPUT delay constant with
varying period
Set OUTPUT trailing delay to 750 ns
TIME
The absolute transition times are held when the
pulse width is varied.
WRATio
The ratio of transition time to pulse width is held
when the pulse width is varied.
:PULS:TRAN:HOLD TIME
:PULS:TRAN:HOLD WRAT
Hold OUTPUT transitions fixed when
pulse width varies
Hold OUTPUT transition width ratio
when pulse width varies