Measurement Operations & Settings Signal Conditioning
160 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:STATe
syntax: :CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:STATe<wsp>OFF(0) | ON(1)
description: Specifies whether the attenuator uses its λ offset table to compensate for wavelength de-
pendent losses in the the test set-up. This table contains, for each wavelength specified, the
additional power offset to be applied.
This command does not affect the module’s internal enviromental temperature and optical
wavelength compensation, which remain active.
parameters: OFF or 0
ON or 1
The offset table is not used to compensate for wavelength dependent losses.
The attenuator adds the appropriate value from its λ offset table to the global
power offset.
response: none
example: CONF1:OFFS:WAV:STAT ON
affects: All attenuator modules.
command:
:CONFigure[n][:CHANnel[m]]:OFFSet:WAVelength:STATe?
syntax: :CONFigure[n]][:CHANnel[m]]:OFFSet:WAVelength:STATe?
description: Queries whether the attenuator uses power values from its λ offset table .
parameters: none
response: boolean 0 The offset table is not used.
1 The attenuator uses its λ offset table.
example: CONF1:OFFS:WAV:STAT? 0<END>
affects: All attenuator modules.

Table of Contents

Other manuals for Agilent Technologies 8166B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8166B and is the answer not in the manual?

Agilent Technologies 8166B Specifications

General IconGeneral
BrandAgilent Technologies
Model8166B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals