The Status Model Introduction to Programming
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 41
Other Commands
*OPT? returns the installed modules and the slots these modules are installed
in:
For example, *OPT? 81682A, 81533B, 81532A, ,
Modules 81682A, 81533B, and 81532A are installed in slots 0 to 2 re-
spectively. Slots 3 and 4 are empty.
*WAI prevents the instrument from executing any further commands until the
current command has finished executing. All pending operations are
completed during the wait period.
*IDN? identifies the instrument; returns the manufacturer, instrument model
number, serial number, and firmware revision level.