EasyManuals Logo

Agilent Technologies 8166B Programming Guide

Agilent Technologies 8166B
274 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
Message Queues Introduction to Programming
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 21
The Output Queue
The output queue contains responses to query messages. The instrument
transmits any data from the output queue when a controller addresses the
instrument as a talker.
Each response message ends with a carriage return (CR, 0D
16
) and a LF
(0A
16
), with EOI=TRUE. If no query is received, or if the query has an error,
the output queue remains empty.
The Message Available bit (MAV, bit 4) is set in the Status Byte register
whenever there is data in the output queue.

Table of Contents

Other manuals for Agilent Technologies 8166B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8166B and is the answer not in the manual?

Agilent Technologies 8166B Specifications

General IconGeneral
BrandAgilent Technologies
Model8166B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals