Triggering - The TRIGger Subsystem Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 175
command:
:TRIGger[n][:CHANnel[m]]:INPut:REARm?
syntax: :TRIGger[n][:CHANnel[m]]:INPut:REARm?
description: Returns the arming response of a channel to an incoming trigger.
parameters: none
response: A boolean value: 0: trigger rearming disabled
1: trigger rearming enabled (default)
example: trig1:inp:rearm? 0<END>
affects: All Agilent 8163A/B series power meter modules, and Agilent 8161x series return loss mod-
ules.
dual sensors: Can only be sent to master channel, slave channel parameters are identical.
command:
:TRIGger[n][:CHANnel[m]]:OFFSet
syntax: :TRIGger[n][:CHANnel[m]]:OFFSet <value>
description: Sets the number of incoming triggers received before data logging begins.
parameters: <value> - an integer value. (maximum possible value is 1e+9)
response: none
example: trig1:offs 5
affects: All Agilent 81636B and 81637B series power meter modules.
command:
:TRIGger[n][:CHANnel[m]]:OFFSet?
syntax: :TRIGger[n][:CHANnel[m]]:OFFSet?
description: Returns the number of incoming triggers received before data logging begins.
parameters: none
response: an integer value.
example: trig1:offs? 5<END>
affects: All Agilent 81636B and 81637B series power meter modules.

Table of Contents

Other manuals for Agilent Technologies 8166B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8166B and is the answer not in the manual?

Agilent Technologies 8166B Specifications

General IconGeneral
BrandAgilent Technologies
Model8166B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals