Status Reporting – The STATus Subsystem Instrument Setup and Status
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 75
command:
:STATusn:QUEStionable:ENABle
syntax: :STATusn:QUEStionable:ENABle<wsp><value>
description: Sets the bits in the Questionable Slot Status Enable Mask (QSSEM) for slot n that enable the
contents of the Questionable Slot Status Register (QSSR) for slot n to affect the QSESR.
Setting a bit in this register to 1 enables the corresponding bit in the QSSER for slot n to af-
fect bit n of the QSESR.
parameters: The bit value for the QSSEM as a 16-bit signed integer value (0 .. +32767)
response: none
example: stat:ques:enab 128
command:
:STATusn:QUEStionable:ENABle?
syntax: :STATusn:QUEStionable:ENABle?
description: Returns the QSSEM for slot n
parameters: none
response: The bit value for the QSSEM as a 16-bit signed integer value (0 .. +32767)
example: stat:ques:enab? → +128<END>