ILT.2156
Replace W67 (cable between A8J8 to A10J1) and go to
the Troubleshooting Foldout, block 1 (located just prior to
the \Special Cases" tab).
ILT.2240
The most common problem with RF power is loose or
damaged connectors and coax cables (especially hard
lines on the RF deck). If the instrument has sudden drops
in power above 10 GHz, check for faulty connections.
Check all hard lines, making sure that none of the
cables are cross-threaded, canted, or damaged. Check
all connections from the YO (A26) to the RF OUTPUT
connector (J1).
Are all connections and cables good?
Go To
ILT.2260 Continue
ILT.2241
Does the instrument have low band (
2 GHz), and does
the RF power problem exist only in low band?
Go To
ILT.2245 Continue
1a-36 Instrument Level Troubleshooting Agilent 8360