EasyManuals Logo

Agilent Technologies 83640B Troubleshooting Guide

Agilent Technologies 83640B
670 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
ILT.2152
Perform the following setup:
1. Reconnect all cables.
2. Set the power switch to standby.
3. Disconnect W67 from A8J8.
4. Connect an oscilloscope to the A8J8 connector.
5. Connect a function generator to the AM input on the
front (or rear) panel.
6. Turn on synthesizer power.
7. Set the function generator for a 0.25 V rms sine wave
at 1.0 MHz.
8. Press
4
PRESET
5
.
The oscilloscope should display the same signal as the
function generator, a sine wave approximately 0.25 V rms
(.707 V p-p) at 1.0 MHz.
Go To
A8.9000
Go To
ILT.2156
ILT.2155
Replace W7 (front or rear panel AM input to A8J7) and
go to the Troubleshooting Foldout, block 1 (located just
prior to the \Special Cases" tab).
Agilent 8360 Instrument Level Troubleshooting 1a-35

Other manuals for Agilent Technologies 83640B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83640B and is the answer not in the manual?

Agilent Technologies 83640B Specifications

General IconGeneral
BrandAgilent Technologies
Model83640B
CategoryPortable Generator
LanguageEnglish

Related product manuals