ILT.2152
Perform the following setup:
1. Reconnect all cables.
2. Set the power switch to standby.
3. Disconnect W67 from A8J8.
4. Connect an oscilloscope to the A8J8 connector.
5. Connect a function generator to the AM input on the
front (or rear) panel.
6. Turn on synthesizer power.
7. Set the function generator for a 0.25 V rms sine wave
at 1.0 MHz.
8. Press
4
PRESET
5
.
The oscilloscope should display the same signal as the
function generator, a sine wave approximately 0.25 V rms
(.707 V p-p) at 1.0 MHz.
Go To
A8.9000
Go To
ILT.2156
ILT.2155
Replace W7 (front or rear panel AM input to A8J7) and
go to the Troubleshooting Foldout, block 1 (located just
prior to the \Special Cases" tab).
Agilent 8360 Instrument Level Troubleshooting 1a-35