ILT.3010
If the display is dim and ickering, check the dim display
option: Press SYSTEM
4
MENU
5
, and select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Dim Display
.
Repeatedly select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Dim Display
to toggle this attribute.
Does the problem remain when you toggle
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Dim Display
?
Done Continue
ILT.3011
The remaining display problems are most likely one of the
following:
Cosmetic fault. The display appearance is fuzzy, varies
in brightness from segment to segment, or another
supercial characteristic is questionable. Replace the
display (go to A21.9000).
The display takes a long time to update
, and UNLOCK
ashes in the display.
Press
4
PRESET
54
SERVICE
5
and select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Unlock Info
. The
following is a prioritized list. Find the rst one that
applies.
1. REF UNLOCK: go to A7.9000.
2. SAMPLER UNLOCK: go to A6.9000.
3. FRACN UNLOCK: go to A4.9000.
4. YO UNLOCK: consider the following, in the order
listed:
a. A6: go to A6.9000.
b. A5: go to A5.9000.
c. A13: go to A13.9000.
d. A26: go to RF.2690.
5. HET UNLOCK: go to RF.2490.
If the display updates slowly, examine (and modify,if
required) calibration constant #554 using the following
procedure:
Press
4
PRESET
54
SERVICE
5
.
Agilent 8360 Instrument Level Troubleshooting 1a-51