Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Adjust Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Calib Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select Cal
.
Press
4
554
54
ENTER
5
.
Calibration constant #554 should be set to 245. If it is
not, continue with the following:
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Modify Cal
.
Press
4
245
54
ENTER
5
.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Cal Util Menu
NNNNNNNNNNNNNNNNNNNNNNNNNN
Save Cal
NNNNNNNNNNN
yes
.
The display has an intermittent fault such as going
blank, or unintelligible characters displayed. Possible
failures include the following:
1. Display (A21).
2. Front panel processor (A3).
3. Cables (W1, W3, W4).
If these alternatives do not x the problem, use
the instrument block diagram (located behind the
\Troubleshooting Block Diagrams" tab) to troubleshoot
this problem.
1a-52 Instrument Level Troubleshooting Agilent 8360