ILT.3030
Check the continuity between front/rear panel AM input
and the end of W7 (connects to A10J1).
Continue
Go To
A10.9000
ILT.3031
Replace the cable (W7) connecting the front/rear panel
AM input to A10J1, and go to the Troubleshooting
Foldout, block 1 (located just prior to the \Special Cases"
tab).
ILT.3040
Check the continuity from the front/rear panel to the end
of W8 (connects to A11J1).
Continue
Go To
A11.9000
1a-54 Instrument Level Troubleshooting Agilent 8360