ILT.3041
Replace the cable (W8) that connects the front/rear
panel FM input to A11J1, and go to the Troubleshooting
Foldout, block 1 (located just prior to the \Special Cases"
tab).
ILT.3050
Check the continuity from the front/rear panel pulse
input and the end of W6 (connects to A9J4).
Continue
Go To
ILT.3052
ILT.3051
Replace the cable (W6) that connects the front/rear
panel pulse input to A9J4 and go to the Troubleshooting
Foldout, block 1 (located just prior to the \Special Cases"
tab).
ILT.3052
Does the pulse problem exist
only
below 2 GHz?
Continue
Go To
ILT.3054
Agilent 8360 Instrument Level Troubleshooting 1a-55