ILT.3053
Does the pulse problem exist
only
above 2 GHz?
Go To
A9.9000
Go To
ILT.3057
ILT.3054
Perform the following key sequence:
1. Press
4
PRESET
54
SERVICE
5
.
2.
Select
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Selftest Menu
.
3. Press
4
108
54
ENTER
5
.
4.
Select
NNNNNNNNNNN
Log
(asterisk on).
5.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Output Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
When to Log Always
.
6. Press
4
PRIOR
5
.
7.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Do Test #108
.
8.
Select
NNNNNNNNNNNNNN
Cont
NNNNNNNNNNNNNN
Cont
.
Go To
A9.9000 Continue
1a-56 Instrument Level Troubleshooting Agilent 8360