ILT.3055
Continue
Go To
RF.2490
ILT.3056
Replace W33 and then go to the Troubleshooting F
oldout,
block 1 (located just prior to the \Special Cases" tab).
ILT.3057
Perform the following key sequence:
1. Press
4
PRESET
54
SERVICE
5
.
2.
Select
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Selftest Menu
.
3. Press
4
108
54
ENTER
5
.
4.
Select
NNNNNNNNNNN
Log
(asterisk on).
5.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Output Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
When to Log Always
.
6. Press
4
PRIOR
5
.
7.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Do Test #108
.
8.
Select
NNNNNNNNNNNNNN
Cont
until
HB Mod OFF [mV]
is displayed.
Make the following measurements.
Agilent 8360 Instrument Level Troubleshooting 1a-57