ILT.3061
Replace W32 and then go to the Troubleshooting Foldout,
block 1 (located just prior to the \Special Cases" tab).
ILT.3062
Check the continuity from the front panel external
ALC input (rear panel for option 004) to the end of W9
(connects to A10J4).
Continue
Go To
A10.9000
ILT.3063
Replace the cable (W9) that connects the front
panel external ALC input to A10J4, and go to the
Troubleshooting Foldout, block 1 (located just prior to the
\Special Cases" tab).
ILT.3070
Using either an oscilloscope or a voltmeter, check the
following voltages at the front panel SMI connector.
Location Measurement
Pin 7 +5.2 Vdc
Pin 5 +15.0 Vdc
Pin 10, 18, 19, 20 TTL High