Did all voltages pass?
Go To
A2.9000 Continue
ILT.3071
Perform the following setup:
1. Press
4
PRESET
5
.
2. Press
4
SWEEP TIME
54
10.0
54
SEC
5
.
Using either an oscilloscope or a voltmeter, check the
following voltages on the front panel SMI connector.
Note that pin 12 is a sweep ramp.To adjust the sweep
ramp, change the instrument sweep time (if you use an
oscilloscope, try a sweep time of approximately 100 ms).
Location Measurement
Pin 15
0
15 Vdc
Pin 6 +8.0 Vdc
Pin 12 Sweep Ramp (0 to 10.0 V)
Did all voltages pass?
Go To
ILT.3075 Continue
Agilent 8360 Instrument Level Troubleshooting 1a-61