ILT.3072
Measure on the center pin at each end of the cable.
Continue
Go To
ILT.3074
ILT.3073
Replace the cable (W10) that connects the front
panel SMI input to the A10 assembly, and go to the
Troubleshooting Foldout, block 1 (located just prior to the
\Special Cases" tab).
ILT.3074
Although the problem has not been isolated, the
following are possible:
The front panel SMI (A2) is faulty.
W2 is faulty (connects A2 to the motherboard).
The ALC assembly (A10) is faulty. This is true if
leveling is the problem.
The source module used with the synthesizer is faulty.
To conrm this, either use the rear panel SMI, or
substitute a new source module or synthesizer.
A remote possibility exists that the motherboard (A22)
is faulty.
If these alternatives do not x the problem, use
the instrument block diagram (located behind the
\Troubleshooting Block Diagrams" tab) to troubleshoot
this problem.
1a-62 Instrument Level Troubleshooting Agilent 8360