ILT.3111
There does not seem to be a failure. The rear panel
output should be 10 MHz at approximately 0 dBm. If a
problem still exists, refer to the instrument block diagram
(located behind the \Troubleshooting Block Diagrams"
tab) to troubleshoot this problem.
ILT.3112
Go To
A7.9000 Continue
ILT.3113
Replace W19, the coaxial cable that connects A7J5 to the
rear panel, and go to the Troubleshooting Foldout, block
1 (located just prior to the \Special Cases" tab).
1a-68 Instrument Level Troubleshooting Agilent 8360