ILT.3120
Before continuing, be sure that auto sweep operates
correctly. If it does not, restart troubleshooting at
ILT.000.
Measure the resistance between the TRIGGER INPUT and
XA15J1-61.
Is the resistance
90 and
110 ?
Go To
ILT.3140 Continue
ILT.3121
Perform the following setup:
1. Press
4
PRESET
5
SWEEP
4
MENU
5
.
2.
Select
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Start Sweep Trigger Ext
.
3. Press
4
SWEEP TIME
54
5
54
SEC
5
.
Go To
A15.9000 Continue
Agilent 8360 Instrument Level Troubleshooting 1a-69