EasyManuals Logo

Agilent Technologies 8564EC User Manual

Agilent Technologies 8564EC
716 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #386 background imageLoading...
Page #386 background image
386 Chapter7
Language Reference
ACPCOMPUTE Adjacent Channel Power Compute
to obtain a valid measurement.
ERR 910 SPAN>ACP indicates that the frequency span is too wide,
compared to the channel bandwidth, to obtain an accurate
measurement.
If any of these errors occurs, the measurement is not completed. To
make a measurement, adjust your instrument state settings depending
on the error that has occurred.
Example
10 OUTPUT 718;"ACPCOMPUTE;"

Table of Contents

Other manuals for Agilent Technologies 8564EC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8564EC and is the answer not in the manual?

Agilent Technologies 8564EC Specifications

General IconGeneral
BrandAgilent Technologies
Model8564EC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals