Chapter 7 417
Language Reference
AMPCORDATA Amplitude Correction Data
AMPCORDATA Amplitude Correction Data
Syntax
Figure 7-51 AMPCORDATA Syntax
Description
The AMPCORDATA function allows you to enter or query the
frequency-amplitude correction points that are used to normalize the
spectrum analyzer measurement. Up to 200 pairs of frequency-
amplitude correction points can be entered.
Whenever ampcor is on, the correction values are added to all
measurement results. Setting ampcor off, performing an instrument
preset, (IP) or turning off the spectrum analyzer turns off the amplitude
correction. Turning ampcor off does not erase the current
frequency-amplitude correction factors.