18-10
Measure Commands
AMPLitude:RNSValue
Restrictions 86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Example 10 OUTPUT 707;”:MEASURE:AMPLITUDE:RNSTABILIZE ON”
Query MEASure:AMPLitude:RNSTabilize?
Returned Format [:MEASure:AMPLitude:RNSTabilize] {1 | 0}<NL>
AMPLitude:RNSValue
Command MEASure:AMPLitude:RNSValue {ONE | ZERO},<RN_value>
Sets the RN stabilization value for the specified signal level.
Restrictions 86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Example 10 OUTPUT 707;”:MEASURE:AMPLITUDE:RNSVALUE ONE,1E-5”
Query MEASure:AMPLitude:RNSValue?
Returned Format [:MEASure:AMPLitude:RNSValue] {ONE | ZERO},<RN_value><NL>
AMPLitude:SAMPlitude?
Query MEASure:AMPLitude:SAMPlitude?
Returns the signal amplitude measurement. Always uses the vertical units of the channel.
Restrictions 86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Example 10 OUTPUT 707;”:MEASURE:AMPLITUDE:SAMPLITUDE?”
Returned Format [:MEASure:AMPLitude:SAMPlitude] <value><NL>
AMPLitude:TI?
Query MEASure:AMPLitude:TI? {ONE | ZERO}
Returns the total interference measurement for the specified logic level. Uses the current
noise and interference units.
Restrictions 86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Example 10 OUTPUT 707;”:MEASURE:AMPLITUDE:TI? ONE”
Returned Format [:MEASure:AMPLitude:TI] <value><NL>
AMPLitude:TI:DEFine
Command MEASure:AMPLitude:TI:DEFine <ber_level>
Specifies the Bit Error Ratio (BER) at which the total interference (TI) is measured. The default value
is 10
-12
. The value is shared with the total jitter (TJ) measurement. Refer to “JITTer:TJ:DEFine” on
page 18-34.
Restrictions 86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Example 10 OUTPUT 707;”:MEASURE:AMPLITUDE:TI:DEFINE -50”