6-21
Error Messages
Error Messages in Alphabetical Order
PRINTER: power off
Error Number
174
The power to the printer at the parallel port is off.
PRINTER: reset in progress
Information
Message
If the printer takes longer than a half-second to reset, this message will be
displayed until printer is finished with reset.
PRINT/PLOT IN PROGRESS, ABORT WITH LOCAL
Error Number
166
If a print or plot is in progress and you attempt a second print or plot, this
message is displayed and the second attempt is ignored. To abort a print or plot
in progress, press .
PROBE POWER SHUT DOWN!
Error Number
23
One or both of the probe power supplies have been shut down due to an
over-current, over-voltage, or under-voltage condition.
PROCESSING DISPLAY LIST
Information
Message
The display information is being processed for a screen print to a copy device
and stored in the copy spool buffer. During this time, the analyzer's resources
are dedicated to this task (which takes less than a few seconds).
PWR MTR NOT ON/CONNECTED OR WRONG ADDRS
Error Number
117
The power meter cannot be accessed by the analyzer. Verify that the power
meter address and model number set in the analyzer match the address and
model number of the actual power meter.
RANGE CAUSED POWER LVL CHANGE IN LIST
Error Number
213
The selected power range changed the power level of one or more segments in
the swept list table. Change the segment power or change the power range.
REQUESTED DATA NOT CURRENTLY AVAILABLE
Error Number
30
The analyzer does not currently contain the data you have requested. For
example, this condition occurs when you request error term arrays and no
calibration is active.
RIPPLE LIMIT TABLE EMPTY.
Information
Message
The ripple limit table does not have any frequency bands defined. Add at least
one frequency band to the ripple limit table for ripple testing.
Local