8753ET Specifications and Characteristics
Instrument Specifications
Table 2-13 Test Port Input
8753ET Test Port Input
Dynamic Accuracy (Characteristic)
For the transmission port, accuracy of the test port input power reading relative to the reference
input power level.
• Input: transmission port
• For test port powers > −60 dBm and < −10 dBm, magnitude dynamic accuracy is
0.02 dB + 0.001 dB/dB from the reference power, phase dynamic accuracy is
0.132 deg + 0.0066 deg/dB from the reference power.