Agilent Flowmeter ADM1000
Operating Instructions 7
Recalibration and Repair Service
To have your ADM1000 recalibrated or repaired, consult the
Agilent website for more information. Go to
www.chem.agilent.com and do a quick search for ADM
Flowmeters. Once there, look in the Technical Support section
for the recalibration/repair details.
Replacing Flexible Tubing
When replacing the flexible tubing please order P/N 701-0016.
This tubing is available by the foot; a minimum of 2 feet is nec-
essary. If tubing other than this is used, please cut to 18 inches
2 inches) in length. Important: If flows to be measured are
less than 200 mL/min and a greater length of tubing is desired
(greater than 18 inches) please order tubing extension kit
P/N 220-1179.
If tubing greater in length than 18 inches is used without this
kit, errors in the displayed flow value may occur with flows
under 200 mL/min.
Product Specifications
Flow range: 0.5 to 1000 mL/min, autoranging
Accuracy ± 3% of reading, or ± 0.2 mL/min, whichever is
greater.
Tubing temperature range: -62°C to 110°C
Operating temperature range: 0°C to 45°C
Power: 9V battery (alkaline)
Display: 16 character alphanumeric
Automatic power off
Split ratio mode with continuous split flow rate reading
NIST traceability
Compatible with non-corrosive and non-flammable gases
Self-checking power on sequence
Each flowmeter individually computer calibrated.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ADM1000 and is the answer not in the manual?

Agilent Technologies ADM1000 Specifications

General IconGeneral
BrandAgilent Technologies
ModelADM1000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals