Chapter 4
Front-Panel Key and SCPI Command Reference
Front-Panel Keys
127
Front-Panel Key and SCPI Command
Reference
4.2.1.3 Low Pass Filter
Sets the type of low pass filters used by the instrument. The frequency indicated is the high frequency cut-off.
4.2.1.4 Band Pass Filter
Selects the band pass filter. This function is only available when PSA firmware revision A.11.08 or above, or
Option 23B is installed.
Mode: MRECEIVE
Key Path:
Det/Demod
Remote Command: [:SENSe]:MRECeive:LPFilter OFF|F3K|F15K|F30K|F300K
[:SENSe]:MRECeive:LPFilter?
Preset: OFF
State Saved: Saved in instrument state.
Range: None (>300 kHz) | 3 kHz | 15 kHz | 30 kHz | 300 kHz
Dependencies/Couplings: For Audio measurements, the 300 kHz choice is disabled, because the cutoff
frequency of FIR LP filter in FPGA of Audio HW is 275 kHz.
For Audio measurements the None (>300 kHz) key will be displayed as None.
Remote Command Notes: Global to the current mode.
Example: MREC:LPF F3K
MREC:LPF?
Mode: MRECEIVE
Key Path:
Det/Demod
Remote Command: [:SENSe]:MRECeive:BPFilter OFF|CCITt
[:SENSe]:MRECeive:BPFilter?
Preset: OFF
State Saved: Saved in instrument state.
Range: None | CCITT
Dependencies/Couplings: None.
Remote Command Notes: Global to the current mode.
Example: MREC:BPF CCIT
MREC:BPF?