Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
193
Front-Panel Key and SCPI Command
Reference
4.4.4.1 SPAN/X Scale
The SPAN/X Scale key accesses the menu to set the desired horizon scale and associated settings of waveform.
4.4.4.1.1 Scale/Div
This key is for Scale/Div control
Table 4-14 AM Depth Numeric Results Window
Name Corresponding Results
AM Depth n=1 1st
AM depth output
Modulation Rate n=1 3rd
Modulation Distortion n=1 4th
Modulation SINAD n=1 5th
Mode: MRECEIVE
Key Path: Span X Scale
Remote Command: :DISPlay:AMDepth[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]:PDIVis
ion <seconds>
:DISPlay:AMDepth[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]:PDIVis
ion?
Unit: Seconds
Preset: 100μs
State Saved: Saved in instrument state.
Min: 1.0 ns
Max: 1 s
Dependencies/Couplings: See Restriction and Notes
Notes: If the Scale Coupling is set to On, this value is automatically determined by the
measurement result. When you set this value manually, Scale Coupling automatically
changes to Off.
Example: :DISP:AMD:WIND:TRAC:X:PDIV 100μs
:DISP:AMD:WIND:TRAC:X:PDIV?

Table of Contents