Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
209
Front-Panel Key and SCPI Command
Reference
4.4.5.1 SPAN/X Scale
The SPAN/X Scale key accesses the menu to set the desired horizontal scale and associated settings of the
waveform.You must have the Demod Waveform selected under the Trace/View key.
4.4.5.1.1 Scale/Div
This key is for Scale/Div control
Table 4-18 FM Deviation Numeric Results Window
Name Corresponding Results
FM Deviation N=1 1st
FM Deviation output
Modulation Rate n=1 3rd
Modulation Distortion n=1 4th
Modulation SINAD n=1 5th
Mode: MRECEIVE
Key Path:
Span X Scale
Remote Command: :DISPlay:FMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]
:PDIVision <seconds>
:DISPlay:FMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]
:PDIVision?
Unit: Seconds
Preset: 100us
State Saved: Saved in instrument state.
Min: –10 s
Max: 1 s
Dependencies/Couplings: See Restriction and Notes
Notes: If the Scale Coupling is set to On, this value is automatically determined by the
measurement result. When you set this value manually, Scale Coupling
automatically changes to Off.
Example: :DISP:FMD:WIND:TRAC:X:PDIV 100us
:DISP:FMD:WIND:TRAC:X:PDIV?