Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
254
Front-Panel Key and SCPI Command
Reference
4.4.10.1.2 Display Unit
Sets unit for Normal display mode.
4.4.10.1.3 Ratio Reference
If Frequency is to be displayed relative to a reference, enter the value as a ratio reference using this key.
Example: :DISP:AUDF:MODE NORM
:DISP:AUDF:MODE?
Mode: MRECEIVE
Key Path:
Amplitude/Y Scale
Remote Command: :DISPlay:AUDFreq:UNIT HZ|KHZ
:DISPlay:AUDFreq:UNIT?
Preset: HZ
State Saved: Saved in instrument state.
Range: Hz|kHz
Example: :DISP:AUDF:UNIT KHZ
:DISP:AUDF:UNIT?
Mode: MRECEIVE
Key Path:
Amplitude/Y Scale
Remote Command: :DISPlay:AUDFreq:RREFerence <freq>
:DISPlay:AUDFreq:RREFerence?
:DISPlay:AUDFreq:RREFerence:AUTO 0|1|OFF|ON
:DISPlay:AUDFreq:RREFerence:AUTO?
Unit: Hz|kHz
Preset: 400 Hz
State Saved: Saved in instrument state.
Min: 20 Hz
Max: 500 kHz
Dependencies/Couplings: No