Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
271
Front-Panel Key and SCPI Command
Reference
4.4.14.1.2 Display Unit
Sets the unit for Normal display.
4.4.14.1.3 Ratio Reference
If the Tuned RF Level is to be displayed relative to a reference, enter the value as a ratio reference using this key.
Example: DISP:TTRF:MODE NORM
DISP:TTRF:MODE?
Mode: MRECEIVE
Key Path:
Amplitude / Y Scale
Remote Command: :DISPlay:TTRFlevel:UNIT DBM|WATT|V|DBMV|DBUV
:DISPlay:TTRFlevel:UNIT?
Preset: DBM
State Saved: Saved in instrument state.
Range: dBm|Watt|V
Example: DISP:TTRF:UNIT DBM
DISP:TTRF:UNIT?
Mode: MRECEIVE
Key Path:
Amplitude / Y Scale
Remote Command: :DISPlay:TTRFlevel:RREFerence <real>
:DISPlay:TTRFlevel:RREFerence?
:DISPlay:TTRFlevel:RREFerence:AUTO 0|1|OFF|ON
:DISPlay:TTRFlevel:RREFerence:AUTO?
Preset: 0dBm
State Saved: Saved in instrument state.
Unit: dBm
Min: –150dBm
Max: 30dBm
Example: DISP:TTRF:RREF 0.0
DISP:TTRF:RREF?
:DISP:TTRF:RREF:AUTO OFF
:DISP:TTRF:RREF:AUTO?