370 Chapter 8
N5531S Measuring Receiver Performance Verification
Performance Verification Tests
N5531S Measuring Receiver Performance
Verification
Table 8-5 Example Test Record
Reference Attenuator Cal Data
S/N
Atten
Calibration
Value
(dB)
Atten
Uncert
(± dB)
Attenuator
Measured
(dB)
Linearity Error
(dB)
(Attenuator Measured
Atten Cal value)
Specification
(± dB)
0000 0
9.989 0.006 9.983 0.006 0.014
20.042 0.006 20.044 0.002 0.019
30.025 0.006 30.022 0.003 0.024
40.194 0.006 40.190 0.004 0.029
50.175 0.011 50.169 0.006 0.034