Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
222
Front-Panel Key and SCPI Command
Reference
4.4.6 PM Deviation Measurement Definition
The PM Deviation measurement consists of 2 views: Numeric Results View and Waveform View.
1. PM Deviation Numeric Results View has two windows:
• PM Deviation Numeric Results Window (upper)
• PM Deviation Settings Window (lower)
2. PM Deviation Demod Waveform View has two windows:
• PM Deviation Waveform Window (upper)
• PM Deviation Numeric Results Window (lower)
Table 4-19 PM Deviation Numeric Results Window
Name Corresponding Results
PM Deviation n=1 1st
PM Deviation output
Modulation Rate n=1 3rd
Modulation Distortion n=1 4th
Modulation SINAD n=1 5th
Table 4-20 PM Deviation Settings Window
Name Corresponding Results
IF BW Type It can be Auto, Manual
IF BW This value is the IF bandwidth set manually
Detector Selection It can be Peak +, Peak -, Peak +/– 2 or RMS
Peak Hold It can be On or Off
HP Filter It can be None, 50 Hz or 300 Hz
LP Filter It can be None, 3 kHz, 15 kHz, 30 kHz or 300 kHz
Table 4-21 PM Deviation Waveform Window
Marker Trace Yes
Corresponding Trace PM Deviation vs. Seconds trace (n=0)

Table of Contents

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals