4.4.6.1 SPAN/X Scale
The SPAN/X Scale key accesses the menu to set the desired horizon scale and associated settings of waveform.
You must have the Demod Waveform selected under the
Trace/View key.
4.4.6.1.1 Scale/Div
This key is for Scale/Div control.
Table 4-22 PM Deviation Numeric Results Window
Name Corresponding Results
PM Deviation n=1 1st
PM Deviation output
Modulation Rate n=1 3rd
Modulation Distortion n=1 4th
Modulation SINAD n=1 5th
Key Path:
Span X Scale
Remote Command: :DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]
:PDIVision <seconds>
:DISPlay:PMDeviation[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]
:PDIVision?
Unit: Seconds
Preset: 100 μs
State Saved: Saved in instrument state.
Min: 1.0 ns
Max: 1 s
Dependencies/Couplings: See Restriction and Notes
Notes: If the Scale Coupling is set to On, this value is automatically determined by the
measurement result. When you set this value manually, Scale Coupling
automatically changes to Off.
Example: :DISP:PMD:WIND:TRAC:X:PDIV 100us
:DISP:PMD:WIND:TRAC:X:PDIV?

Table of Contents

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals