Add Option 1E1 (Attenuator)
Kit Part Numbers: E8251-60384, E8251-60385, and E8251-60383
Introduction
This upgrade kit adds Option 1E1 (Attenuator) to the E8257D (Analog) PSG Signal Generators.
Installation includes the following major steps, starting on page 6:
1. Check signal generator functionality, page 6.
2. Remove the outer and inner signal generator covers, starting on page 7.
3. Install Option 1E1, page 10.
4. Activate Option 1E1, page 13.
5. Verify proper installation of Option 1E1, page 16
6. Re-assemble the signal generator, page 16.
7. Verify signal generator calibration, page 16.
Product Affected: . . . . . . . . . . . . . . . . . . . . . . . . . E8257D
Serial Numbers: . . . . . . . . . . . . . . . . . . . . . . . . . . All
Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
To Be Performed By: . . . . . . . . . . . . . . . . . . . . . . (X) Agilent Technologies Service Center
(X) Personnel Qualified by Agilent Technologies
(X) Customer
Estimated Installation Time: . . . . . . . . . . . . . . . 1.5 hours
Estimated Verification Time: . . . . . . . . . . . . . . . Option 520 = 8 hours
Option 540 = 15 hours
Option 550 = 16 hours
Option 567 = 16 hours