Chapter 3 19
Operation Verification
Performing a Self-Test
Self-Test Failure
If a self-test failure occurs, perform the following procedure:
1. Make sure all external cables, including GPIB, LAN, and RS-232 cables, are disconnected from the
signal generator and perform the self-test again.
2. If the self-test continues to fail, send the signal generator to an Agilent service center for repair, with a
detailed description of the failed test(s) and any other error messages that appeared on the display.
See “Returning a Signal Generator to Agilent Technologies” on page 16 for return instructions.