3-18 Service Guide E8364-90038
Tests and Adjustments PNA Series Microwave Network Analyzers
Preliminary Checks E8362C, E8363C, E836C
Figure 3-7 Typical Cable Connector Repeatability Response
If the Cable Connector Repeatability Check Fails
1. Clean the cable and devices, and torque connections to specifications (refer to “Review the Principles of
Connector Care” on page 3-5). Repeat the check.
2. If the check still fails, the cable should be repaired or replaced.