Chapter 2 51
Front-Panel Key Reference
AMPLITUDE Y Scale
AMPLITUDE Y Scale
Activates the reference level function and accesses the amplitude menu
keys. Amplitude menu keys allow you to set functions that affect the
way data on the vertical axis is displayed or corrected.
Ref Level Allows the reference level to be changed. This function is activated
when
AMPLITUDE Y Scale is pressed. The reference level is the
amplitude power or voltage represented by the top graticule on the
display. Changing the value of the reference level changes the absolute
amplitude level (in the selected amplitude units) of the top graticule
line. The reference level can be changed using the step keys, the knob,
or the numeric keypad. Pressing any digit (0 through 9) on the numeric
keypad brings up the terminator menu.
Key Access:
AMPLITUDE / Y Scale
Attenuation
Auto Man
Sets the input attenuation to manual or automatic and allows you to set
the attenuation level in 5 dB increments when
Attenuation (Man) is
selected. The analyzer input attenuator, which is normally coupled to
the reference level control, reduces the power level of the analyzer input
signal at the input mixer. The attenuator is recoupled when Attenuation
(Auto) is selected. Attenuation can be changed using the step keys, the
knob, or the numeric keypad. To select 0 dB, the numeric keypad must
be used (except for E7401A.)
Key Access:
AMPLITUDE / Y Scale
CAUTION To prevent damage to the input mixer, do not exceed a power level of
+30 dBm at the input. To prevent signal compression, keep the power at
the input mixer below 0 dBm. With the attenuator set to Auto, a signal
at or below the reference level will result in a mixer level at or below
the Max Mixer Lvl.
Scale/Div Sets the logarithmic units per vertical graticule division on the display.
The
Scale/Div function is only available when the Scale Type key is set to
Log. Scale/Div values may range from 0.1 to 20 dB per division With
Demod View (On) selected, the range is 1 kHz to 240 kHz/div.
Key Access:
AMPLITUDE / Y Scale
Scale Type
Log Lin
Scales the vertical graticule divisions in logarithmic units when Log is
underlined. Logarithmic units may range from 0.1 to 20 dB per
division. When Lin is underlined, the vertical divisions are linearly
scaled and the default amplitude units are volts. The top of the display
is set to the reference-level value and the bottom graticule becomes zero
volts. (Each division of the graticule is one-tenth of the reference level
in volts.) Pressing
Scale Type always sets the units specified for the

Other manuals for Agilent Technologies EMC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EMC Series and is the answer not in the manual?

Agilent Technologies EMC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelEMC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals