Chapter 2 67
Front-Panel Key Reference
Auto Couple
The Average detector displays the average of the signal within the
bucket. The averaging method depends upon
Avg Type selection
(Video or RMS).
The
Peak detector displays the maximum of the signal within the
bucket.
The
Sample detector displays the instantaneous level of the signal at
the time of the sampling.
The
Negative Peak detector displays the minimum of the signal within
the bucket.
Neither average nor sample detectors measure amplitudes of CW
signals as accurately as peak, because they may not find a spectral
component’s true peak, but they do measure noise without the biases of
peak detection.
The detector in use is indicated on the left side of the display. A # will
appear next to it if the detector has been manually selected.
Key Access:
Auto Couple
Auto
The system selects peak detection as the default, but if
a condition arises where a different type of detection
scheme provides greater accuracy, the system will use
the alternate scheme. For example, when using the
Marker Noise function, Auto mode selects Average
detection for Res BWs > 300 Hz and Sample for
Res BWs ≤ 300 Hz as shown in Figure 2-1.
Key Access:
Auto Couple, Detector Auto Man