EasyManua.ls Logo

Agilent Technologies ENA Series

Agilent Technologies ENA Series
9 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Product Declassification and Security Product Name: ENA series RF Network Analyzer
Model Number(s): E5071C
Page 4 of 9
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the size of
memory, how it is used, its location, volatility, and the sanitization procedure.
Summary of instrument memory - base instrument
Memory
Type and
Size
Writable During
Normal Operation?
Data Retained When
Powered Off?
Purpose/Contents
Data Input
Method
Location in
Instrument and
Remarks
Sanitization
Procedure
Main
Memory
(RAM)
1GB
Yes
No
Windows Operating
system memory
Operating
system
(not user)
Digital Mother Board
Cycle power
Media
Storage
(Hard Disk
Drive)
40 GB
Yes
Yes
Windows Operating
system boot device,
factory correction
data, service log,
and users file
including saved
traces data, settings,
or images.
User-saved
data
HDD assembly in the
instrument (Option
019) or
Removable Hard
Disk Drive (Option
017)
Remove
Memory for
PCI and
DSP
(RAM)
2.5MB
Yes
No
Data Processing for
measurement
Measurement
(not user)
PCI DSP Card
Cycle power
Non-volatile
memory
(Flash)
1MB
No
Yes
Product serial
number, options,
correction constants,
offsets, DAC values
Adjustment
program
performed by
Agilent factory
personnel or
by calibration
labs
PCI DSP Card
N/A
(The data is
not stored by
user under
normal
operation. )

Other manuals for Agilent Technologies ENA Series

Related product manuals