Installation Note E4407-90041
5
Torque Settings
To avoid potential RFI leakage and prevent connector damage, tighten to the following torque limits:
WARNING Before you disassemble the instrument, turn the power switch to Standby and unplug the instrument.
Failure to unplug the instrument can result in personal injury.
CAUTION Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic
assemblies should be performed at a static-safe workstation. Refer to the documentation that pertains to your
instrument for information about static-safe workstations and ordering static-safe accessories.
Item
Torque
In-lb N-cm
SMA Connectors 8.5 95
3-mm, T-10 TORX screws 9102
4-mm, T-15 TORX screws 21 236