Agilent 1200 Series Valves 33
Maintenance, Repair, and Troubleshooting 3
Maintenance and Repair
Resetting the EMF counter
1 Switch to the Diagnostics View of the ChemStation.
2 Select Valve from the Maintenance menu.
3 In the upcoming dialog box click on Reset Counter (see also Figure 17 on
page 37).
The procedure to reset the EMF counter with the Agilent 1200 Series control
module is described on page 23.