EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G2260A

Agilent Technologies G2260A User Manual

Agilent Technologies G2260A
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #103 background imageLoading...
Page #103 background image
103
Maintenance
7
Simple Repairs
3
Rotate the gripper arm slowly from left to right and apply a
gentle pressure to the paper clip. The clip will engage on
an internal catch and the rotation of the arm will be
blocked.
4 Hold the paper clip in place, press the gripper release
button and rotate the gripper arm to the right.
5 The gripper arm will come off. 6 Replace the gripper arm by holding the paper clip in place,
pushing the gripper arm into the holder and rotating the
gripper arm to the left.
Next Steps:
7 On completion of this procedure: Install the front cover.
8 Turn the power to the autosampler ON.

Table of Contents

Other manuals for Agilent Technologies G2260A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G2260A and is the answer not in the manual?

Agilent Technologies G2260A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG2260A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals