EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G2260A

Agilent Technologies G2260A User Manual

Agilent Technologies G2260A
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
11
Introduction to the Autosampler
1
Sampling Sequence
Injection Sequence
Before the start of the injection sequence, and during an analysis, the injection
valve is in the mainpass position (Figure 2 on page 11). In this position, the
mobile phase flows through the autosamplers metering device, sample loop,
and needle, ensuring all parts in contact with sample are flushed during the
run, thus minimizing carry-over.
Figure 2 Mainpass Position
When the sample sequence begins, the valve unit switches to the bypass
position (Figure 3 on page 11). Solvent from the pump enters the valve unit at
port 1, and flows directly to the column through port 6.
Figure 3 Bypass Position

Table of Contents

Other manuals for Agilent Technologies G2260A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G2260A and is the answer not in the manual?

Agilent Technologies G2260A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG2260A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals