EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G2260A

Agilent Technologies G2260A User Manual

Agilent Technologies G2260A
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
64
5 Optimizing Performance
Optimization for Lowest Carry-over
Using an Injector Program
The process is based on a program that switches the bypass grove of the
injection valve into the flow line for cleaning. This switching event is
performed at the end of the equilibration time to ensure that the bypass grove
is filled with the start concentration of the mobile phase. Otherwise the
separation could be influenced, especially if microbore columns are used.
For example:
Outside wash of needle in vial 7 before injection
Injector program:
Draw x.x (y) µl from sample
NEEDLE WASH vial 7
Inject
Wait (equilibration time - see text above)
Valve bypass
Wait 0.2 min
Valve mainpass
Valve bypass
Valve mainpass
NOTE
Overlapped injection together with additional injection valve switching is not possible.

Table of Contents

Other manuals for Agilent Technologies G2260A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G2260A and is the answer not in the manual?

Agilent Technologies G2260A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG2260A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals