Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 113
7 Error Information
MCT Error Messages
MCT Error Messages
These errors are module specific.
VALVE INIT FAILED
Error ID: 32030
During initialization, the valve actuator turns until the encoder reads the reference
index position. An error is generated, if the reference index cannot be found
within a given time.
Probable cause Suggested actions
1 Defect in cable connection of valve drive con-
trol.
Please contact your Agilent service representa-
tive.
2 Defect in cable connection of valve actuator
encoder reader.
Please contact your Agilent service representa-
tive.
3 Defect of valve drive or valve actuator
encoder reader.
Please contact your Agilent service representa-
tive.