Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 46
3 Using the Module
Status Indicators
Status Indicators
This procedure exemplarily shows an arbitrary LC stack configuration.
1 The module status indicator indicates one of six possible module conditions:
Status indicators
1. Idle
2. Run mode
3. Not-ready. Waiting for a specific pre-run condition to be reached or
completed.
4. Error mode - interrupts the analysis and requires attention (for example, a
leak or defective internal components).
5. Resident mode (blinking) - for example, during update of main firmware.
6. Bootloader mode (fast blinking). Try to re-boot the module or try a
cold-start. Then try a firmware update.